Acclarix AX4 Diagnostic Ultrasound System User Manual
Maintenance
 Inspect the monitor and reusable accessories after they are cleaned and disinfected.
The console (including touch screen, track pad and hard keys) of system is splash resistant (IPX 1)
but not waterproof. Do not immerse or expose to extended moisture. Splash resistance does not
extend to transducer connectors. Keep connectors dry.

12.4.1.Cleaning the System Surface

The validated disinfecting agents for cleaning the system are:
Ethanol (75%)
Isopropanol (70%)
To clean the system surface:
1.
Switch off the system and unplug it.
2.
Use a soft cloth dampened with disinfecting agent to gently wipe the entire exterior surface,
including the screen, of the equipment thoroughly until no visible contaminants remain.
3.
Wipe off the disinfecting agent with a fresh cloth or towel, dampened with tap water after
cleaning until no visible cleaning agent remains.
4.
Dry the system in a ventilated and cool place.
NOTE:
1.
Make sure the system is free of gel and any other visible residue.
2.
Use a soft dry cloth without chemicals for cleaning. The monitor surface is easily
scratched.

12.4.2.Needle Guide Bracket Cleaning and Sterilization

NOTE:
1. Use proper sterilization technique at all times when performing a biopsy.
2. Ensure that protective gloves are not worn.
1. The needle guide bracket kits are not disinfected or sterilized before delivery. The
operators must clean and sterilize the needle guide bracket kits before the first use
and after each use.
Inspect the bracket for damage such as cracks or breakage. If damage is evident,
2.
discontinue use of bracket and contact your Edan representative for disposal
guidance.
3. Sterilize the bracket before disposal or sending back to manufacturer for repair.
Cleaning
12.4.2.1.
1. Wear sterile protective gloves to prevent infection.
2. Disconnect the needle guide bracket from the transducer after each use, and remove all visible
WARNING
WARNING
- 132 -