Step 2: Select the correct size guide for the gauge of needle to be used, and secure it to the needle
guide holder by tightening knob 1. If the needle guide holder is loose tighten the screw with an
allen wrench. Loosen knob 2 all the way prior to installing the bracket on the transducer.
Step 3: The transducer has a small circular dimple on one side of the transducer, and a rectangular
ridge on the other side. Place the needle guide bracket over the transducer such that knob 2
tightens into the circular dimple and the ridge sits inside the corresponding slot on the bracket.
(see figures 7-13 through 7-16).
Figure 7-13 Convex slot of Probe
Screw
螺钉
矩形卡槽
Rectangular slot
Figure 7-12Structure of Needle Guide Bracket Kit
Convex
一字形凸起
圆锥台
Cone
Figure 7-14 Notch of Probe
- 91 -
Features
旋钮2
Knob 2
Notch
凹槽