Table of Contents
loading

Summary of Contents for EDAN SE-101

  • Page 2 This manual will help you understand the operation and maintenance of the product better. It is reminded that the product shall be used strictly complying with this manual. User’s operation failing to comply with this manual may result in malfunction or accident for which EDAN INSTRUMENTS, INC. (hereinafter called EDAN) can not be held liable.
  • Page 3 Terms Used in this Manual This guide is designed to give key concepts on safety precautions. WARNING A WARNING label advises against certain actions or situations that could result in personal injury or death. CAUTION A CAUTION label advises against actions or situations that could damage equipment, produce inaccurate data, or invalidate a procedure.
  • Page 4: Table Of Contents

    Table of Contents Chapter 1 Safety Guidance ......................1 1.1 Intended Use........................1 1.2 Warnings and Cautions ......................1 1.2.1 General Warnings ...................... 1 1.2.2 Battery Care Warnings ....................4 1.2.3 General Cautions ....................... 5 1.2.4 Operation for Wireless System.................. 5 1.2.5 Preparation and Operation Warnings (for Exercise ECG) ........
  • Page 5 5.6 Inspection Before Test ....................... 37 5.7 Setting DX12 Transmitter (for Wireless System) ............. 38 5.7.1 Keyboard Locking/Unlocking ................. 40 5.7.2 Menu Settings ......................41 Chapter 6 Operation Instructions for Resting ECG ..............42 6.1 Viewing Lead Placement Information ................43 6.2 Selecting a Patient Record to Start a New Test ..............
  • Page 6 6.6.5 Analyzing HRV ....................... 72 6.6.5.1 Editing the HRV Data on the Analysis Screen ..........73 6.6.5.2 Editing the HRV Waveform in the Waveform Window ........ 74 6.6.5.3 Previewing HRV ................... 75 6.6.6 Analyzing HRT......................76 6.6.7 Analyzing Vector ECG .................... 78 6.6.7.1 Displaying Vector ECG with All Plane and All Loop ........
  • Page 7 7.7 Recovery Phase ....................... 100 7.8 Exiting the Exercise Test ....................100 7.9 About Analysis Screen ....................101 7.9.1 About Summary Screen..................101 7.9.2 About ST Analysis Screen ..................103 7.9.3 About All View Review Screen ................105 7.9.4 About Trend Screen ....................107 7.9.5 About ECG Strip Screen ..................
  • Page 8 9.2.6 Setting Anti-aliasing ....................125 9.2.7 Selecting QRS Voice ..................... 125 9.2.8 Selecting Sequence Mode When Sampling ............125 9.3 Device Setup ........................126 9.3.1 Setting Sampling Device ..................126 9.3.2 Setting Device Type/Mode ..................127 9.3.3 Setting COM Port of Sample/Treadmill/BP Monitor ..........127 9.3.4 Setting a Protocol ....................
  • Page 9 Chapter 10 Hint Information ....................146 Chapter 11 Cleaning, Care and Maintenance ................. 148 11.1 Cleaning and Maintaining the Treadmill ............... 148 11.2 Cleaning and Maintaining the Patient Cable and Reusable Electrodes ......148 11.3 Disinfection ........................149 11.4 Maintenance of ECG Sampling Box ................149 Chapter 12 Accessories ......................
  • Page 10: Chapter 1 Safety Guidance

    SE-1010 PC ECG User Manual Safety Guidance Chapter 1 Safety Guidance This chapter provides important safety information related to the use of SE-1010 PC ECG. 1.1 Intended Use SE-1010 PC ECG is a PC-based diagnostic tool intended to acquire, process and store ECG signals from adult and pediatric patients undergoing stress exercise test or resting test.
  • Page 11 SE-1010 PC ECG User Manual Safety Guidance WARNING 3. EXPLOSION HAZARD - Do not use the system in the presence of flammable anesthetic mixtures with oxygen or other flammable agents. 4. SHOCK HAZARD - The power receptacle must be a hospital grade grounded outlet. Never try to adapt the three-prong plug to fit a two-slot outlet.
  • Page 12 SE-1010 PC ECG User Manual Safety Guidance WARNING 16. SHOCK HAZARD - Don’t connect electrical equipment, which has not been supplied as a part of the system, to the multiple portable socket-outlet supplying the system. 17. Do not connect any equipment or accessories that are not approved by the manufacturer or that are not IEC/EN 60601-1-1 approved to the system.
  • Page 13: Battery Care Warnings

    SE-1010 PC ECG User Manual Safety Guidance WARNING 25. You should purchase computer, printer, treadmill, ergometer, BP monitor and bar code reader from the manufacturer. Otherwise, the manufacturer will not be held responsible for the maintenance of the PC hardware, operating system and other accessories.
  • Page 14: General Cautions

    SE-1010 PC ECG User Manual Safety Guidance 1.2.3 General Cautions CAUTION 1. Avoid liquid splash and excessive temperature. The temperature must be kept between 5 ºC and 40 ºC during operation, and it should be kept between -20 ºC and 55 ºC during transportation and storage.
  • Page 15: Preparation And Operation Warnings (For Exercise Ecg)

    SE-1010 PC ECG User Manual Safety Guidance 1.2.5 Preparation and Operation Warnings (for Exercise ECG) WARNING 1. Test the safety stop (mushroom type) and safety stop (cord type) of the treadmill before using the system. 2. During the exercise test, ensure that there are at least 2 experienced physicians present.
  • Page 16: Contraindications (For Exercise Ecg)

    SE-1010 PC ECG User Manual Safety Guidance 1.2.6 Contraindications (for Exercise ECG) Absolute Contraindications: 1. Acute MI (within 2 days) 2. High-risk unstable angina 3. Hemodynamic compromise caused by uncontrolled cardiac arrhythmia 4. Symptomatic severe aortic stenosis 5. Heart failure with clinic episode uncontrolled 6.
  • Page 17: List Of Symbols

    SE-1010 PC ECG User Manual Safety Guidance 1.3 List of Symbols Equipment or part of CF type with defibrillator proof Caution Consult Instructions for Use Recycle Part Number Serial Number Date of Manufacture Manufacturer Authorized Representative in the European Community The symbol indicates that the device complies with the European Council Directive 93/42/EEC concerning medical devices.
  • Page 18 SE-1010 PC ECG User Manual Safety Guidance Class Ⅱ Transmission Status Indicator of Bluetooth Burglar Lock Power Supply Indicator of DX12 Receiver This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (for DX12 Transmitter) (1) this device may not cause harmful interference, and (2) this device must accept any interference received,...
  • Page 19: Chapter 2 Introduction

    SE-1010 PC ECG User Manual Introduction Chapter 2 Introduction SE-1010 PC ECG has similar functions with an ordinary electrocardiograph. ECG data can be sampled, analyzed and stored in a PC, and it can be saved in PDF, Word, BMP or JPG format. ECG waves can be frozen and reviewed.
  • Page 20 SE-1010 PC ECG User Manual Introduction Wired System of SE-1010 PC ECG System 1. Resting ECG of Wired System Patient Patient Cable Resting PC (Manually ECG Cable DP12 ECG Configured) Sampling Box USB Cable Printer (Manually Configured) 2. Exercise ECG of Wired System Treadmill or Ergometer (Manually Configured) Patient...
  • Page 21 SE-1010 PC ECG User Manual Introduction Wireless System of SE-1010 PC ECG System The DX12 device which consists of transmitter and receiver has passed FCC certification. This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
  • Page 22: Dp12 Ecg Sampling Box Of Wired System

    SE-1010 PC ECG User Manual Introduction FCC Statement This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation.
  • Page 23 SE-1010 PC ECG User Manual Introduction Front Panel Lamp USB Socket Name Explanation When the ECG sampling box is powered by the PC, the lamp Lamp will be lit. USB Socket Connecting to the USB socket of the PC with a USB cable - 14 -...
  • Page 24 SE-1010 PC ECG User Manual Introduction USB Socket Definitions of corresponding pins: Signal Signal WARNING 1. When the computer connected to the USB cable is powered on, do not connect the USB cable to the DP12 ECG sampling box; when the system is powered on, do not disconnect the USB cable from the ECG sampling box.
  • Page 25 SE-1010 PC ECG User Manual Introduction Patient Cable Socket Definitions of corresponding pins: Signal Signal Signal C2 / V2 F / LL C3 / V3 C1 / V1 or NC C4 / V4 C1 / V1 C5 / V5 R / RA RF (N) /RL or NC C6 / V6...
  • Page 26: Dx12 Ecg Sampling Boxes Of Wireless System

    SE-1010 PC ECG User Manual Introduction WARNING 1. Accessory equipment connected to the analog and digital interfaces must be certified according to the respective IEC/EN standards (e.g. IEC/EN 60950 for data processing equipment and IEC/EN 60601-1 for medical equipment). Furthermore all configuration shall comply with the valid version of the standard IEC/EN 60601-1-1.
  • Page 27 SE-1010 PC ECG User Manual Introduction Front Panel Patient Cable Socket Patient Cable Socket :Applied part of type CF with defibrillator proof :Caution Definitions of corresponding pins: Signal Signal Signal Signal C5/V5 L/LA F/LL C2/V2 C4/V4 R/RA C6/V6 C1/V1 C3/V3 N/RL NOTE: The left side of “/”...
  • Page 28 SE-1010 PC ECG User Manual Introduction DX12 Receiver Appearance USB Socket USB Socket - 19 -...
  • Page 29: Features

    SE-1010 PC ECG User Manual Introduction Definitions of corresponding pins: Signal Signal Lock 2.4 Features 1. Powerful functions, friendly windows and easy operation 2. 3/6/12-channel ECG waves are displayed and printed simultaneously 3. ECG waves can be frozen and reviewed 4.
  • Page 30 SE-1010 PC ECG User Manual Introduction 7. Reports can be printed in PDF, Word, JPG or BMP format 8. Supporting multi-language 9. Updated to be a network electrocardiograph, transmitting ECG data over LAN or WAN or INTERNET 10. Automatic baseline adjustment for optimal printing 11.
  • Page 31: Chapter 3 Assembling Se-1010 Pc Ecg System

    SE-1010 PC ECG User Manual Assembling SE-1010 PC ECG System Chapter 3 Assembling SE-1010 PC ECG System 3.1 Assembling Wired System Patient Cable for Resting ECG Patient Cable for Exercise ECG DP12 ECG Sampling Box Exercise ECG Cable Resting ECG Cable Assembly Drawing - 22 -...
  • Page 32 SE-1010 PC ECG User Manual Assembling SE-1010 PC ECG System For Resting ECG of wired system, ♦ Insert plug 1 of the patient cable into socket 3 of DP12 ECG sampling box. ♦ Insert plug 8 of the cable into socket 4 of DP12 ECG sampling box. ♦...
  • Page 33: Assembling Wireless System

    SE-1010 PC ECG User Manual Assembling SE-1010 PC ECG System 3.2 Assembling Wireless System Patient Cable Patient Cable DX12 Transmitter  D X12 Belt Resting ECG Cable - 24 -...
  • Page 34 SE-1010 PC ECG User Manual Assembling SE-1010 PC ECG System DX12 Receiver Exercise ECG Cable Burglar Lock Assembly Drawing For Resting ECG of wireless system, 1. Insert plug 1 of the patient cable into socket 2 of DX12 transmitter. 2. Insert plug 4 of the cable into socket 6 of DX12 receiver. 3.
  • Page 35 SE-1010 PC ECG User Manual Assembling SE-1010 PC ECG System 4. Insert plug 9 of the cable into the USB socket of the PC. 5. Connect plug 10 of the cable to the BP monitor. 6. Connect a treadmill or an ergometer to the PC. 7.
  • Page 36: Chapter 4 Installing Se-1010 Pc Ecg Software

    SE-1010 PC ECG User Manual Installing SE-1010 PC ECG Software Chapter 4 Installing SE-1010 PC ECG Software 4.1 System Running Environment 4.1.1 Requirements on the Hardware of the PC CPU: Pentium P4, Celeron D 310 or above System Memory (RAM): 512MB or above Main Board Recommend the main board of Intel chipset...
  • Page 37: About Installation Window

    SE-1010 PC ECG User Manual Installing SE-1010 PC ECG Software 4.2 About Installation Window Insert installation into CD-ROM, double-click Setup.exe to open the following installation window. Figure 4-1 Installation Window Click on the Install button to install PC ECG. Click on the Next button continually during installation.
  • Page 38: Chapter 5 Preparations Before Operation

    SE-1010 PC ECG User Manual Preparations Before Operation Chapter 5 Preparations Before Operation 5.1 Preparing the Patient 5.1.1 Instructing the Patient Before attaching the electrodes, greet the patient and explain the procedure. Explaining the procedure decreases the patient’s anxiety. Reassure the patient that the procedure is painless. Privacy is important for relaxation.
  • Page 39: Connecting The Electrodes Of Wired System

    SE-1010 PC ECG User Manual Preparations Before Operation 5.2 Connecting the Electrodes of Wired System WARNING The performance and electric shock protection can be guaranteed only if the original patient cable and electrodes of the manufacturer are used. The patient cable includes main cable and lead wires which can be connected to electrodes according to the colors and identifiers.
  • Page 40: Connecting The Electrodes Of Wireless System

    SE-1010 PC ECG User Manual Preparations Before Operation 5.3 Connecting the Electrodes of Wireless System The patient cable includes main cable and lead wires which can be connected to electrodes according to the colors and identifiers. Lead Wires Main Cable Connecting to DX12 Electrode Connector transmitter...
  • Page 41: Wilson Lead System

    SE-1010 PC ECG User Manual Preparations Before Operation Table 5-1 Electrodes and Their Identifiers and Color Codes European American WILSON FRANK Identifier Color Code Identifier Color Code Right arm Right arm White Left arm Left arm Yellow Black Right leg Right leg N or RF Black...
  • Page 42: Frank Lead System

    SE-1010 PC ECG User Manual Preparations Before Operation 5.4.2 Frank Lead System FRANK lead system is usually adopted when PC ECG is used to produce VCG. The conventional letter designations for the electrodes and their respective positions are: E/C2: at the front midline M/C5: at the back midline I/C1: at the right mid-axillary line A/C4: at the left mid-axillary line...
  • Page 43 SE-1010 PC ECG User Manual Preparations Before Operation 1. Clean the electrode area which is in a short distance above the ankle or the wrist with alcohol. 2. Daub the electrode area on the limb with gel evenly. 3. Place a small amount of gel on the metal part of the limb electrode clamp. 4.
  • Page 44: Attaching Electrodes To The Patient (For Exercise Ecg)

    SE-1010 PC ECG User Manual Preparations Before Operation ♦ Attach the disposable electrode to the electrode area at the back midline. The quality of ECG waveform will be affected by the contact resistance between the patient and the electrode. In order to get a high-quality ECG, the skin-electrode resistance must be minimized while connecting electrodes.
  • Page 45 SE-1010 PC ECG User Manual Preparations Before Operation White/Red Brown/Red White/Yellow Brown/Yellow White/Green Brown/Green White/Brown Brown/Blue White/Black Brown/Orange White/Violet Brown/Violet The Precordial Electrodes’ Positions on Body Surface: C1: Fourth intercostal space at the right border of the sternum C2: Fourth intercostal space at the left border of the sternum C3: Fifth rib between C2 and C4 C4: Fifth intercostal space on the left midclavicular line C5: Left anterior axillary line at the horizontal level of C4...
  • Page 46: Inspection Before Test

    SE-1010 PC ECG User Manual Preparations Before Operation 5.6 Inspection Before Test In order to avoid safety hazards and get good ECG records, the following inspection procedure is recommended before operation. 1. Environment: 1. Make sure that there is no electromagnetic interference source around the equipment, especially large medical electrical equipment such as electrosurgical equipment, radiological equipment, magnetic resonance imaging equipment etc.
  • Page 47: Setting Dx12 Transmitter (For Wireless System)

    SE-1010 PC ECG User Manual Preparations Before Operation 5.7 Setting DX12 Transmitter (for Wireless System) Switch on DX12 receiver and install batteries on DX12 transmitter. Press to start up DX12 transmitter, and then the company information and the main screen will be displayed. Figure 5-1 Main Screen When the main screen is displayed, press to switch the leads.
  • Page 48 SE-1010 PC ECG User Manual Preparations Before Operation 1) Press to enter the menu screen. Figure 5-2 Menu Screen 2) Press to display Match Device item in black, and then press to open the screen with a prompt “Inquiring…”. - 39 -...
  • Page 49: Keyboard Locking/Unlocking

    SE-1010 PC ECG User Manual Preparations Before Operation 3) When the receiver is found, the address of DX12 receiver will be displayed on the screen in 10 seconds. Press to start up the Bluetooth connection. The Bluetooth connection will be displayed on the main screen of DX12 transmitter and the Bluetooth status indicator illuminates in kelly when DX12 transmitter and DX12 receiver are matched successfully.
  • Page 50: Menu Settings

    You can see the related information, such as Software version:1.0 software version, ID, address of DX12 receiver, ID:0016a400035D manufacture and release time about the device. Device Information EDAN NOTE: The device information is for 2010.04.20 reference only. - 41 -...
  • Page 51: Chapter 6 Operation Instructions For Resting Ecg

    SE-1010 PC ECG User Manual Operation Instructions for Resting ECG Chapter 6 Operation Instructions for Resting ECG Double-click on the shortcut icon on the desktop to display the main screen. the desktop icon for SE-1010 PC ECG. NOTE: Do not use other software when using PC ECG software. Figure 6-1 Toolbar of Main Screen The toolbar contains six buttons.
  • Page 52: Viewing Lead Placement Information

    SE-1010 PC ECG User Manual Operation Instructions for Resting ECG You can set the unit, print, frequency of AC and the saving path of source data based on your needs. Click on the OK button after setup, the system will enter the main screen automatically. NOTE: You should install the software to the saving path of source data after the uninstallation and reinstallation;...
  • Page 53: Selecting A Patient Record To Start A New Test

    SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.2 Selecting a Patient Record to Start a New Test Click on the Data Manager button on the main screen (Figure 6-1) to open the Data Manager screen (Figure 6-3). Figure 6-3 Data Manager Screen 1.
  • Page 54 SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 3. Or, click on Advanced Search to display the Search Condition window. Enter the search conditions, and click on the Search button, and all the patient records which meet the conditions will be displayed in the patient information list.
  • Page 55: Entering New Patient Information

    SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.3 Entering New Patient Information 6.3.1 Entering New Patient Information Manually If the patient is a new one, you can click on the New Patient button on the main screen (Figure 6-1) to display the Patient Information window. Then you need to input the patient’s related information.
  • Page 56 SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 3. Enter information of doctor and department 1) Enter information of physician, technician or Req. physician Click on the pull-down list , and then click on the Edit button to display the Edit window.
  • Page 57 SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 2) Enter information of department Click on the pull-down list , and then click on the Edit button to display the Edit Department window. Enter the department name in the textbox, and then click on the Add button. The doctor name will be displayed in List.
  • Page 58: Entering Patient Information By Using A Bar Code Reader

    SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 4. Confirm pacemaker information If you select Pacemaker in the Print Setting window, Pacemaker appears in the Patient Information window. Select Pacemaker to detect very small pacemaker pulses. However, when pacemaker pulse enhancer is on, the system is very sensitive, and should not be close to equipment emitting high frequency radiation.
  • Page 59: Selecting Sampling Type

    SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.4 Selecting Sampling Type You can select a sampling type in the Patient information window. 6.5 Sampling Resting ECG After inputting the patient information, click on the OK button in the Patient information window to open the ECG sampling screen.
  • Page 60: Specifying Display Mode

    SE-1010 PC ECG User Manual Operation Instructions for Resting ECG The system begins to pre-sample ECG. Figure 6-5 Pre-Sampling Screen 6.5.1 Specifying Display Mode There are eight display modes including 12*1, 6*2, 3*4, 6*1, 3*1, 3*4+1, 3*4+3 and 6*2+1. When the display mode is set to 12*1, 12-channel ECG waves are displayed on one screen simultaneously.
  • Page 61: Specifying Lowpass Filter

    SE-1010 PC ECG User Manual Operation Instructions for Resting ECG When the display mode is set to 3*1, 3-channel ECG waves are displayed on one screen simultaneously. When the display mode is set to 3*4+1, 12-channel ECG waves are displayed in 4 groups of 3 and one rhythm lead on one screen.
  • Page 62: Recording Ecg Data

    SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.5.5 Recording ECG Data When the pre-sample ECG waves are steady, you can click on the Start button to save the sampled ECG data to the designated directory. For details, please refer Section 9.1.4, “Specifying the Storage Path of the ECG Data”.
  • Page 63: Stopping Sampling Data

    SE-1010 PC ECG User Manual Operation Instructions for Resting ECG Figure 6-7 Wave Review Window Click on Exit to return to the ECG sampling screen. NOTE: The display modes in the Wave review window are the modes you select on the ECG sampling screen.
  • Page 64: Printing Ecg Waves

    SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.5.8 Printing ECG Waves Click on the Print button on the ECG sampling screen to print the ECG waves on the ECG sampling screen. NOTE: 1. You can set the printer type on the Printer and Faxes screen. For details, please refer to Section 6.6.10, “Printing ECG Reports”.
  • Page 65: Viewing The Waveform

    SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.1.1 Viewing the Waveform Click on the Waveform tab on the normal ECG analysis screen to open the Waveform window (Figure 6-8). Figure 6-8 Normal ECG - Waveform Window You can choose a speed, a gain and a display mode for the displayed waves. Click on the Measure button in the Waveform window (Figure 6-8).
  • Page 66 SE-1010 PC ECG User Manual Operation Instructions for Resting ECG The common parameters are displayed in the circle in Figure 6-8. These parameters are as follows. Designation Description Heart Rate Heart Rate P Duration P-wave duration of the current lead PR Dur.
  • Page 67: About The Average Template Window

    SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 1. Enter your own opinions in the Diagnosis textbox, and then click on the Save button. 2. Or, double-click on the necessary results required to be added in the Glossary textbox, the selected results will be displayed in the Diagnosis textbox, and then click on the Save button.
  • Page 68 SE-1010 PC ECG User Manual Operation Instructions for Resting ECG To Edit Waveform on the Analysis Screen You can press a lead button in display magnified average templates of this lead. When you press more than one lead button, magnified average templates of these leads will be overlapped with the same central axis. When you press ALL, magnified average templates of all leads will be overlapped with the same central axis.
  • Page 69: About The Detail Information Window

    SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.1.3 About the Detail Information Window Click on the Detail information tab on the normal ECG analysis screen to open the Detail information window. This window displays lead parameter values as Figure 6-10 shows. Figure 6-10 Normal ECG - Detail Information Window Click on the Export Excel button to export an Excel file.
  • Page 70: About The Rhythm Wave Window

    SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.1.4 About the Rhythm Wave Window Click on the Rhythm Wave tab on the normal ECG analysis screen to open the Rhythm wave window. This window displays rhythm wave as Figure 6-11 shows. Figure 6-11 Normal ECG - Rhythm Wave Window You can set the gain, the speed and the lead of the displayed ECG waves.
  • Page 71 SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 4. Click on Close to close the normal ECG preview screen and return to the previous screen. Figure 6-12 ECG Wave NOTE: The report title is in one line by default, you can also change the line to two lines. Find the text named PCECG.ini in the installation path, modify LineofTitle=0 to LineofTitle=1, and then save the text to change the line to two lines.
  • Page 72: Analyzing Qt Dispersion

    SE-1010 PC ECG User Manual Operation Instructions for Resting ECG Report Title in Two Lines 6.6.2 Analyzing QT Dispersion Click on the QTD button to open the QT Dispersion screen. QT Dispersion: The difference between the largest QT interval and the shortest QT interval based on the synchronous 12-lead surface ECG.
  • Page 73: Editing Waveform On The Qt Dispersion Screen

    SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.2.1 Editing Waveform on the QT Dispersion Screen There are red and blue triangle icons on top of the displayed ECG waves. You can click on the triangle icon to change the color. represents the current chosen R wave represents the position of R wave You can set the speed and the gain of the displayed ECG waves.
  • Page 74: Previewing Qt Dispersion

    SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.2.3 Previewing QT Dispersion Click on the Preview button on the QT Dispersion screen to open the QT Dispersion preview screen. is the toolbar on the QT Dispersion preview screen. 1.
  • Page 75: Analyzing Frequency Ecg

    SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.3 Analyzing Frequency ECG Click on the FCG button to open the Frequency ECG screen. Frequency ECG displays characteristic waves of ECG signal spectrum. There are two tabs on the Frequency ECG screen: Two leads compare and 12-lead power spectrum.
  • Page 76 SE-1010 PC ECG User Manual Operation Instructions for Resting ECG Every parameter of Frequency ECG is shown in the following table. Designation Definition If the first peak value is lower than 90% of the second peak value, it is positive (‘+’). Otherwise it is minus (‘-’). The first peak is too low or disappears.
  • Page 77: About 12-Lead Power Spectrum Window

    SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.3.2 About 12-lead Power Spectrum Window The 12-lead power spectrum window displays the power spectrum of 12-lead ECG waves. Figure 6-16 Frequency ECG - 12-lead Power Spectrum Window 6.6.3.3 Previewing Frequency ECG Click on the Preview button to open the Frequency ECG preview screen.
  • Page 78 SE-1010 PC ECG User Manual Operation Instructions for Resting ECG Figure 6-17 Two-lead Comparison Report Figure 6-18 12-lead Power Spectrum Report - 69 -...
  • Page 79: Analyzing High Frequency Ecg

    SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.4 Analyzing High Frequency ECG In a conventional electrocardiogram, only these ECG signals, of which the amplitude is in the millivolt range and the frequency is less than 100 Hz, are visible. Those ECG signals, of which the amplitude is in the microvolt range and the frequency is between 150Hz and 250Hz, are invisible.
  • Page 80 SE-1010 PC ECG User Manual Operation Instructions for Resting ECG The ECG wave between the two green lines is the present one. Click on to select another QRS wave. Select the present lead in the pull-down list You can choose the lead group, the gain and the speed of the waves displayed in the bottom part of the HF ECG analysis screen.
  • Page 81: Analyzing Hrv

    SE-1010 PC ECG User Manual Operation Instructions for Resting ECG Figure 6-20 HF ECG Report 6.6.5 Analyzing HRV Click on HRV to display the HRV ECG analysis screen. The HRV ECG analysis screen includes two tabs: Auto diagnosis result and Waveform. NOTE: 1.
  • Page 82: Editing The Hrv Data On The Analysis Screen

    SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.5.1 Editing the HRV Data on the Analysis Screen Figure 6-21 Analysis Screen of HRV Designation Definition Sampling time Set sampling time Total Beat Total beat number during the measuring course Heart Rate Heart rate Average RR interval...
  • Page 83: Editing The Hrv Waveform In The Waveform Window

    SE-1010 PC ECG User Manual Operation Instructions for Resting ECG Low Frequency High Frequency LF/HF Ratio of low frequency to high frequency LF (norm) Standard LF power HF (norm) Standard HF power H- Doctor Diagnosis Field 1. Enter your own opinions in the Auto diagnosis textbox, and then click on the Save button. 2.
  • Page 84: Previewing Hrv

    SE-1010 PC ECG User Manual Operation Instructions for Resting ECG HRV waveform is displayed in the Waveform window (Figure 6-22). 1. You can drag the mouse in the window to choose the wave field to be printed. Then click on the Print button to print the selected wave field.
  • Page 85: Analyzing Hrt

    SE-1010 PC ECG User Manual Operation Instructions for Resting ECG Figure 6-23 HRV Preview Screen 6.6.6 Analyzing HRT Click on HRT to display the HRT ECG analysis screen. - 76 -...
  • Page 86 SE-1010 PC ECG User Manual Operation Instructions for Resting ECG You can select Average Histogram or Single Histogram in Mode View field. Click on Next PVC or Previous PVC to see the Next or Previous PVC wave and the relevant histogram.
  • Page 87: Analyzing Vector Ecg

    SE-1010 PC ECG User Manual Operation Instructions for Resting ECG 6.6.7 Analyzing Vector ECG Click on the VCG button on the ECG analysis screen to display the VCG analysis screen. Vector ECG displays 3D image of ECG activity. You can choose the plane and the vector loop on the VCG analysis screen.
  • Page 88 SE-1010 PC ECG User Manual Operation Instructions for Resting ECG Figure 6-24 displays Vector ECG with the plane of ALL and the loop of ALL. A- Vector ECG of Frontal (F) B- Vector ECG of Horizontal (H) C- Vector ECG of Sagittal (S) D- Average templates of X, Y and Z leads.