Acclarix AX4 Diagnostic Ultrasound System User Manual
Join: This joins the selected network. If a password is needed a dialog box will be displayed
to enter it.
Forget Network: This forgets any password of the selected network that had been entered,
but the network is still shown in the list.
Advanced Settings: This brings up a dialog box with additional settings for the selected
wireless network. The dialog box includes the following:
 Password: This is useful when the network password has changed.
 Method: Either Manually entered (Static) or DHCP. Generally wireless networks
use DHCP, and a Static address is only needed if required by the target PACS system.
 IP Address, Subnet Mask, Gateway, DNS: If you chose DHCP then your network
router will set these automatically. If you chose Manually Entered IP then contact
your network IT manager for the appropriate settings.

9.2.2. DICOM

This screen configures DICOM services, including Store and Modality Worklist.
Local Settings:
These settings determine how the system communicates with other DICOM devices. These are
required for both Store and for Modality Worklist.
Local AE Title: Any 16 characters that uniquely identify this system on your DICOM
network. The default "EDAN_STR_SCU" will work unless you have multiple Edan systems
on your network.
Time Out, Transfer Syntax: These determine how the DICOM messages are configured.
The system default settings are common, and you will typically not have to change them. If
Ping works, but Verify does not, you may need to review the conformance statement of your
server to determine the proper settings for these.
Figure 9-7 DICOM Settings