Adding a new Measure Preset:
A new measure preset can be added by copying an existing preset. The new measure preset
contains the same measurements as the copied preset. Select a preset from the list along the
left, and then press the Copy As button.
Enabling/Disabling a measurement in a Preset
Select the preset you want to configure from the list along the left side of the main screen.
The current contents show in the grid along the right of the main screen.
measurement from the measurement library on the touch screen to toggle it in or out of the
selected preset.
Editing an Existing Measurement in a Measure Preset
1.
Select the preset you want to customize from the list along the left of the main screen.
2.
Select the measurement you want to customize from the grid along the right of the main
screen.
3.
Press the Option button on the touch screen to customize the selected measurement.
4.
If the selected measurement can be performed by different tools (Caliper, Trace, etc.) select
the tool you want from the tabs provided, and check the default box.
5.
Check the boxes for the results that you want to derive from this measurement.
For a Generic measurement, press Default button on the touch screen to configure it as a default
measurement for the current selected image mode in a preset.
Figure 8-13 Measure Preset touch screen