Step 3: Insert the needle into the round hole, as shown in Figure 7-22.
7.5.2. Activating Needle Guide Function
To enable the needle guide function:
1.
In the B mode imaging, press Needle button on touch screen, A needle touch screen UI is
displayed, Press Enable button to active the Needle Guide function.
2.
Press Double Line button to switch double line and single line as the Needle guide Line
graphics.
3.
Some needle guide brackets support multiple (up to 2) angles. If the current transducer
supports such a guide then the Angle button appears. Pressing it cycles through the available
angles.
NOTE:
The distance between each dot of the needle guide line indicates 0.5 cm.
Figure 7-21 Turning the Knob 1 Clockwise
Figure 7-22 Insert the Needle into the Round Hole
Figure 7-23 Needle Guide Touch Screen
Round hole