Chapter 10 Cleaning, Care and Maintenance

10.1 Cleaning

Turn off the tablet PC before cleaning. The mains supply must be switched off when
using the adapter to charge the tablet PC.

10.1.1 Cleaning Tablet PC and the Patient Cable

1. Wipe the external surfaces of the tablet PC and the patient cable with a soft cloth dampened
in any of the approved cleaning solutions listed below.
2. Wring excess moisture from the cloth before cleaning.
Recommended Cleaning Solutions
♦ Mild soap and water
♦ 75% alcohol

10.1.2 Cleaning the Reusable Electrodes

1. Remove the remainder gel from the electrodes with a clean soft cloth first.
2. Take suction bulbs and metal cups of chest electrodes apart, and take clamps and metal parts
of limb electrodes apart.
3. Clean them in warm water and make sure there is no remainder gel.
4. Dry the electrodes with a clean dry cloth or air dry naturally.
1. Prevent the detergent from seeping into the tablet PC while cleaning. Do not immerse
the tablet PC or the patient cable into liquid under any circumstances.
2. Do not clean the unit and accessories with abrasive fabric and avoid scratching the
electrodes.
3. Any remainder of detergent should be removed from the tablet PC and the patient
cable after cleaning.
CAUTION
CAUTION
- 65 -
Cleaning, Care and Maintenance