System Cleaning And Disinfection - EDAN U60 User Manual

Diagnostic ultrasound system
Table of Contents

Advertisement

U60 Diagnostic Ultrasound System User Manual
1. Ensure the bracket is cleaned, sterilized and completely dried before storage.
2. Store the bracket in a sterile environment or in a disposable sterile package.
3. Store the bracket under the following conditions:
a) Atmospheric Temp.: -20℃〜+55℃
b) Relative Humidity: 15%~95% (Non-condensing)
c) Atmospheric Pressure: 70kPa ~ 106kPa.

15.2.3. System Cleaning and Disinfection

System Surface Cleaning
The validated disinfecting agents for cleaning the system are:
Ethanol (75%)
Isopropanol (70%)
To clean the system surface:
1.
Switch off the system and unplug it.
2.
Wear sterile protective gloves to prevent infection.
3.
Remove all residual foreign matters from the system surface using sterile cloth or paper
towel immediately after examination.
4.
Use a sterile cloth dampened with cleaning agent to gently wipe the entire exterior surface,
including the screen, of the equipment thoroughly until no visible contaminants remain.
5.
After cleaning, wipe off the cleaning agent with a sterile cloth dampened with tap water until
no visible cleaning agent remains.
6.
Wipe off with a dry sterile cloth to remove residual moisture.
7.
Leave the system to air dry.
8.
If the system is not visually clean at the end of the cleaning steps, please repeat the cleaning
steps through step 4 to step 7.
9.
Inspect the system to ensure that there is no damage.
NOTE:
1.
Make sure they are free of gel and any other visible residue.
2.
Use a soft dry cloth without chemicals for cleaning, because the surface of polarizer
is very soft and easily scratched.
- 184
-
Inspection and Maintenance