EDAN Acclarix AX4 User Manual page 146

Diagnostic ultrasound system, version 1.6
Table of Contents

Advertisement

Acclarix AX4 Diagnostic Ultrasound System User Manual
Guidance and manufacture's declaration – electromagnetic immunity
The system is intended for use in the electromagnetic environment specified below. The customer or
the user of the system should assure that they are used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC/EN 61000-4-2
Electrical fast
transient/burst
IEC/EN 61000-4-4
Surge
IEC/EN 61000-4-5
Power frequency
(50/60Hz)
magnetic field
IEC/EN 61000-4-8
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC/EN 61000-4-11
NOTE
U
is the a.c. mains voltage prior to application of the test level.
T
Electromagnetic Immunity
IEC/EN 60601 test
level
 8 kV contact
 15 kV air
2 kV for power
supply lines
1 kV for line to
line
2 kV for line to
ground
30 A/m
0 % UT; 0.5 cycle
At 0° , 45° , 90° ,
135° , 180° , 225° ,
270° and 315°
0 % UT; 1 cycle
and
70 % UT; 25/30
cycles )
Single phase: at 0°
0 % UT; 250/300
cycle
Compliance level
 8 kV contact
 15 kV air
2kV for power
supply lines
1 kV for line to line
2 kV for line to
ground
30 A/m
0 % U
0.5 cycle
T;
At 0° , 45° , 90° , 135° ,
180° , 225° , 270° and
315°
0 % U
; 1 cycle
T
and
70
%
U
;
25/30
T
cycles )
Single phase: at 0°
0 % U
; 250/300 cycle
T
- 139 -
EMC Information
Electromagnetic environment
- guidance
Floors should be wood, concrete
or ceramic tile. If floor are
covered with synthetic material,
the relative humidity should be at
least 30%.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the system requires continued
operation during power mains
interruptions, it is recommended
that the system be powered from
an uninterruptible power supply
or a battery.