5.5 M-mode

5.5.1. M-mode Touch Screen Controls

The following touchscreen controls impact the M-mode image. The location of the touchscreen
controls can be customized (see section 3.8 for details).
Name
Control
Colorize
Gray Map
Tint
Dynamic
Range
Focus Pos
Sweep
Speed
Strip Size
Side-by-side
Line
Persistence
Frequency
Description
Switches between grayscale and colorized (pseudo-color)
postprocessing maps.
Adjusts the current postprocessing, either grayscale or tinted.
The Dynamic Range, or log compression, adjusts how echo
intensities are converted to brightness. A high dynamic range
will display more shades of gray, while a low dynamic range will
display fewer shades of gray and a more contrasty M-mode
display.
The focus position set in B-mode applies to M-mode as well.
Upward presses move the focus shallower, regardless of the
U/D invert status of the M-mode. Downward presses move the
focus deeper.
Sweep adjusts the sweep speed of the M-mode strip. Options of
Slow, Low, Med High and
increase sweep speed. Downward presses decrease sweep
speed.
Changes the relative size of the M-mode strip compared to the
reference image. Full, large, Med and small are available.
This is an on/off control. When On, the M-Mode strip is
displayed side-by-side with the B-mode image. When Off, the
M-Mode strip is displayed below the B-mode image
The line Persistence determines how many M-mode lines are
averaged together for the display (similar to persistence in
B-Mode). Off, Low, Med and High settings are available.
Frequency determines the transmit frequency used by M-Mode.
Upward presses increase the frequency. Downward presses
decrease the frequency.