Acclarix AX4 Diagnostic Ultrasound System User Manual
8. Inspect the transducer to ensure that there is no damage.
Disinfectant
Figure 6-2 Depth of the Transducer Immerged into Disinfectant
1. Do not immerse the transducer connector. If the cable connector is immersed, do not
plug the connector into the system. Rinse the connector under running water and dry
it thoroughly. If necessary, contact EDAN for service.
2. Prohibit infiltration of any type of liquid into the device or the transducer.
3. Do not immerse the cable and connector of the transducer into solutions. Transducers
can be submerged to, but not including, the strain relief of the transducer array. Do not
immerse or soak any part of a transducer in any cleaning material not listed in the
recommended list of disinfectant.
4. Use the immersion method to disinfect the intra-cavitary and intra-operative
transducer(E8-4Q and L17-7SQ).
5. Only non-immersion method can be used with solution of ethanol or isopropanol. The
solution concentration should not exceed the value provided above.
6. The immersion time should not exceed the time that is specified by the disinfectant
manufacturer.
7. Patient contact area should be immersed into the solution while using the immersion
method, but should not exceed the depth shown in figure 6-2.
L17-7SQ
WARNING
- 61 -
E8-4Q
Transducer