Setting the Presented Bucket

About Bucket
Positions
Checking the Bucket
Alignment
Selecting the Bucket
to Present to the
Door
You can set bucket 1 or bucket 2 to be presented when the VSpin door
opens.
This bucket position is an example of a teachpoint. For this reason, the
process of setting a bucket position is called teaching.
For more information about teachpoints, see "About Teachpoints" on
page 13.
Before you can set the bucket to present to the door, the bucket edge
must be aligned parallel to the open door edge so that a robot will
accurately place a plate on the bucket.
Accurate bucket alignment is also necessary to allow the rotor to lock
properly.
Check the bucket alignment before you proceed.
00085
VSpin
bucket alignment
If the bucket is not aligned correctly, see "Aligning the Buckets" on
page 40.
To select the bucket that is presented to the door:
1. Click the
tab.
Control
The
page opens.
Control
2. Choose the bucket to present to the door.
For example, if you choose bucket 1:
a. Click the
Bucket 1
box.
b. Click
Go to bucket
The door opens with bucket 1 presented.
If the bucket 1 is already presented to the door, then no change
occurs.
Chapter 4: Performing a Spin
radio button in the
Bucket to Present
in the
group box.
Execute