Agilent Technologies Velocity11 VSpin User Manual page 52

Microplate centrifuge
Table of Contents

Advertisement

42
Chapter 3: Installation
VSpin User Guide
00086
VSpin
adjusting bucket
00086
The following illustration shows the bucket aligned with the top
edge of the door.
00085
VSpin
bucket alignment
4. Click
.
Capture 1
5. Without changing the position of the buckets, click
6. Turn on the air source to the VSpin.
7. Click
.
OK
The new teachpoint is listed on the
Mechanical Alignment
To mechanically align and teach the bucket position:
1. Unlock and open the door.
See "Unlocking and Opening the Door Manually" on page 69.
!! DAMAGE HAZARD !!
the bucket is improperly aligned, turning the rotor manually
could cause the rotor or the bucket to collide with the bucket
lock and damage the bucket, rotor or bucket lock.
2. Disconnect the air supply.
See "Disconnecting the Air Supply" on page 67.
If you do not disconnect the air supply or turn off the air source, the
door will close when the rotor is turned.
3. Insert the robot gripper into the VSpin so that the gripper is centered
in the VSpin door.
The gripper must be inserted far enough that when the rotor is
turned, the buckets contact the gripper
screen.
Profiles
If the door lock is extended when
Bucket
Bucket
.
Capture 2

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents