4. Manually turn the rotor slightly so that the left side of the bucket
lightly touches the gripper.
5. Click
.
Capture 1
6. Turn the rotor slightly so that the right side of the bucket lightly
touches the gripper.
7. Click
.
Capture 2
8. Turn on the air source to the VSpin.
9. Click
.
OK
The new teachpoint is listed on the
10. Remove the gripper from the VSpin door.
If you do not remove the gripper, it obstructs the closing of the door.
.
Chapter 3: Installation
VSpin User Guide
screen.