78
Chapter 5: Maintenance and Troubleshooting
VSpin User Guide
Checking Error
Messages
Door Closed Status
Anomaly
If you have problems with the actuators, error messages are displayed in
popup windows, and logged in the Messages group box.
The following types of error and status messages appear in the Messages
group box of the Diagnostics page:
❑
Bucket lock failed
❑
Door lock failed
❑
Door open failed
If an error message is displayed, see the "VSpin Error Reference" on
page 79.
The Door closed sensor is not triggered when the door is fully closed,
but rather when it is just starting to close.
As a result, it is occasionally possible for the door to be open slightly, but
for the Door closed status light to still be lit.
In this regard, it is perhaps more accurate to think of the Door closed
sensor as a "door not open" sensor.