82
Chapter 5: Maintenance and Troubleshooting
VSpin User Guide
Error
Timed out waiting for bucket to
lock
Timed out waiting for bucket to
unlock
Timed out waiting for door to
close
Timed out waiting for door to
lock
Timed out waiting for door to
open
Timed out waiting for door to
unlock
Timed out waiting for servo to
home
Timed out waiting for access to
COM port
Timed out waiting for amplifier
to reset
Timed out waiting for servo to
door
Tried to actuate bucket lock but
failed
Unrecognized hardware
stepping
Unrecognized hardware
stepping. Assuming hardware
stepping 0
What this means
The action did not complete
within the allotted time.
The power or
communications to the
VSpin has been interrupted.
A hardware problem has
occurred that needs to be
addressed.
The system may have
insufficient air pressure or
another hardware problem may
have occurred.
Occurs after initialize if
hardware is unrecognized.
This error occurs when the
software tries to query the VSpin
or a version number, but
receives no response.
For example, VSpin models
older than version 3 do not
report their version number to
the software.
In this case the software
assumes a stepping value of 0.
What to do
Check to make sure that the:
Power cord is properly
plugged in to the VSpin
power entry and the wall
outlet
Fuse is not burned out
Power switch is turned on
The serial cable is properly
plugged into the VSpin and
the computer
Perform the action again.
If the problem persists, contact
Velocity11.
Check to make sure that the:
Air hose is properly
connected to the VSpin and
to the air source
Air is turned on at the
source
Air pressure is at the proper
level as listed in "Laboratory
Requirements" on page 23
Contact Velocity11.
If you have received this error,
you might have a VSpin older
than version 3.
Contact Velocity11 to resolve
this issue.