Chapter 4: Performing a Spin
VSpin User Guide

Configuring Time Settings

The Time Group Box
Changing the Timer
Mode
Where to Go Next
From the Time group box on the Control page you can set and monitor
the duration of a spin.
Timer mode settings let you configure how the spin time is measured.
To change the timer mode settings:
1. Click one of the following radio buttons:
Setting
Total time
Time at speed
Continuous spin
2. Select a time from the list box, at the lower left corner of the
group box.
This controls the amount of time the VSpin spins in the
and
Time at speed
Before you start a spin, refer to the "Workflow Overview" on page 46.
If you have received an error, see "VSpin Error Reference" on page 79.
Definition
Sets the next spin session to last for the specified
duration, including the time it takes to accelerate
and decelerate.
Sets the next spin session to spin for the specified
duration in the time setting selection box, not
including the time it takes to accelerate or
decelerate.
Sets the next spin session to spin until told to stop.
There is no time setting for this mode.
To stop a spin see "Stopping a Spin" on page 60.
modes.
Time
Total time