Agilent Technologies Velocity11 VSpin User Manual page 91

Microplate centrifuge
Table of Contents

Advertisement

Error
NmcNoOp() failed:
Communication likely failed
Profile does not exist
Sensor failed to indicate bucket
unlock before the attempted
move
Sensor failed to indicate door
lock before the attempted move
Sensor indicated imbalance
before the attempted move
ServoClearBits() failed
ServoLoadTraj() failed
ServoResetPos() failed
ServoSetGain() failed
ServoSetHoming() failed
ServoStopMotor() failed
What this means
The VSpin may have become
disconnected during operation.
The VSpin profile does not exist.
The sensor is not working
properly, the door, bucket or lock
is not in a state that allows a spin
to occur, or there is a bucket
imbalance.
Internal error.
Chapter 5: Maintenance and Troubleshooting
What to do
Check to make sure that the:
Power cord is properly
plugged in to the VSpin
power entry and the wall
outlet
Power switch is turned on
Serial cable is properly
plugged into the VSpin and
the computer
Create a new profile or select an
existing one.
Contact Velocity11.
Contact Velocity11.
VSpin User Guide
81

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents