Status light
Color
Balanced
Green
In motion
Green
Amp enabled
Green
Homing
Yellow
After your spin completes, refer to the "Workflow Overview" on page 46.
Chapter 4: Performing a Spin
Illuminates when...
The plates are balanced.
The VSpin must be spinning to detect an
imbalance.
To balance the buckets, see "Starting a Spin"
on page 58.
The rotor is spinning.
The amplifier is enabled.
This status light is for diagnostics purposes,
and helps a Velocity11 engineer to
troubleshoot problems with your VSpin.
The VSpin motor cannot function unless the
amplifier is enabled.
This status light only appears if your model of
VSpin supports it.
The buckets are returning to their home
position. "About the Home Position" on
page 13.