Mounting The Vspin - Agilent Technologies Velocity11 VSpin User Manual

Microplate centrifuge
Table of Contents

Advertisement

Mounting the VSpin

Required
Preparation
About the Procedure
Mounting the VSpin
This procedure assumes that you have already:
Prepared your mounting surface
Drilled the necessary holes
Tapped the screw threads in your benchtop or mounting plate.
Velocity11 provides four M6 X 30 flathead screws. You may need to
obtain screws that are a more suitable length for your particular
laboratory setup.
!! DAMAGE HAZARD !!
during this procedure, the circuit board will be exposed. This
can be easily damaged. Do not touch the circuit board, and
ground yourself with a device such as an anti-static wrist-band to
avoid static discharge.
!! INJURY HAZARD !!
the covers are off. Touching the circuit board when the power is
on could give you a shock.
To mount the VSpin to a stable surface:
1. Remove the front and back VSpin covers, using the following
diagram for reference.
There are four retention screws and washers for each cover.
Use a 2.5mm allen wrench to remove the screws.
Cover retention screw and washer (x4)
00076
When you remove the VSpin covers
Do not turn on the VSpin power when
Back cover
00076
VSpin
shroud removal
Front cover
Chapter 3: Installation
VSpin User Guide
29