Mains Input Supply Connector - Edwards iXL600 Instruction Manual

Dry pump systems
Table of Contents

Advertisement

5. For the three live conductors (L1, L2, L3), carefully remove the cable insulation to the exact dimension specified
in the appropriate figure. Do not twist the cable strands.
6. Referring to
Figure
10, connect the three live conductors to the following connections on the connector block:
L1 to a1, L2 to a2 and L3 to b1. Insert each wire completely into the contact chamber until the copper strands
reach the bottom. Tighten the connection using a 2mm allen key as shown in
Keep the cable in position while applying the recommended tightening torque.
Control/connector
Item
identification
1
Connector Block
2
L1
3
Cable Gland
4
Hood
7. Fit the earth (ground) wire to the Protective Earth connection on the connector block as shown in
using the following procedure:
Prepare the end of the 6 mm2 earth wire to a strip length of 10 mm and twist the strands.
Insert the wire into the earth terminal block.
Tighten the connection using a flat blade screwdriver.
8. Refit the outer cover to the connector block then tighten the cable gland.
© Edwards Limited 2015. All rights reserved.
Edwards and the Edwards logo are trademarks of Edwards Limited.
Figure 10 - Mains input supply connector
Control/connector
Item
identification
5
3-Phase Mains Cable
6
L3
7
Earth
8
L2
A540-55-880 Issue D
Figure 11
to a torque of 1.1 Nm.
Control/connector
Item
identification
9
b1
10
Protective Earth
11
a2
12
a1
Figure 10
Page 23