Edwards iXL600 Instruction Manual page 64

Dry pump systems
Table of Contents

Advertisement

A540-55-880 Issue D
Warning Message
'Action' Message on
on PDT
PDT
Warning 54.11
See manual
MB Temp High
Warning 54.13
-
Sensor missing
Warning 55.11
See manual
DP Temp High
Warning 55.13
-
Sensor missing
Warning 62.11
See manual
MB E/C Temp High
Warning 62.13
-
Sensor missing
Warning 71.13
See manual
AC Sup Missing
Warning 176.01
xxxx yyyy
MB INV xxxx yyyy
aaaaaaaaaaaaaaaa
Warning 176.13
See manual
No MB Inv Comms
Warning 186.01
xxxx yyyy
DP INV xxxx yyyy
aaaaaaaaaaaaaaaa
Warning 186.13
See manual
No DP Inv Comms
Note:
Some of these warnings apply to accessories and may only be encountered if the particular accessory is
fitted.
Page 56
Table A13 - Warnings (continued)
Possible Cause
The temperature of the booster pump is
too high.
The booster temperature sensor may have
become disconnected or failed.
The internal temperature of the pump is
too high.
The dry pump temperature sensor may
have become disconnected or failed.
The temperature of the booster end cover
is too high.
The booster end cover temperature sensor
may have become disconnected or failed.
The pump controller cannot communicate
with the accessory module
Booster Inverter has raised a warning code. Refer to
The pump controller cannot communicate
with the booster inverter.
Dry pump Inverter has raised a warning
code.
The pump controller cannot communicate
with the dry pump inverter.
Edwards and the Edwards logo are trademarks of Edwards Limited.
Action
- Check that cooling water is
connected, switched on and
is to specification given in
Section
2.6.
- Check that the sensor is
fitted and is correctly
connected.
- Check that cooling water is
connected, switched on and
is to specification given in
Section
2.6.
- Check that the sensor is
fitted and is correctly
connected.
- Check that cooling water is
connected, switched on and
is to specification given in
Section
2.6.
- Check that the sensor is
fitted and is correctly
connected.
- Check that the accessory
module is correctly
connected to the pump and
then cycle the power to the
pump.
- If this warning persists,
contact Edwards.
A2.3
information.
Check the wiring between
the pump controller and the
booster inverter.
Refer to
A2.3
information.
Check the wiring between
the pump controller and the
dry pump inverter.
© Edwards Limited 2015. All rights reserved.
for more
for more