Edwards iXL600 Instruction Manual page 69

Dry pump systems
Table of Contents

Advertisement

Fault Code
BIT
Combination
6
8
HIGHTC
4
HIGHTS
2
HIGHTM
1
MISSING_PHASE
7
8
UTCREG
4
UTSREG
2
UTMREG
1
LVREG
8
8
LOWVCC
4
LOWTC
2
LOWTS
1
LOWTM
Example:
PDT shows the following:
0108 0200
This translates to:
alarm 1 in bit 2 – PWM_TRIP
alarm 8 in bit 4 – OVERT
warning 2 in bit 6 - HIGHTM.
This means the motor overheated and switched the output off.
© Edwards Limited 2015. All rights reserved.
Edwards and the Edwards logo are trademarks of Edwards Limited.
Table A17 - Inverter warnings codes (continued)
Inverter warnings
Name
Controller temperature above upper limit / Controller
temperature sensor open circuit - check water supply
Inverter heat sink temperature above upper limit /
Inverter heat sink temperature sensor open circuit - check
water supply
Motor temperature above upper limit / Motor temperature
sensor open circuit - check water supply
The loss of one of the three input phases has been
detected - check mains supply to drive or blown fuses
Upper controller temperature regulator active - current
limit reduced - check cooling
Upper heat sink temperature regulator active - current
limit reduced - check cooling
Upper motor temperature regulator active - check motor
cooling
Lower voltage regulator active - check mains voltage
Internal power supply fault - contact Edwards
Low controller temperature - contact Edwards
Low heatsink temperature - contact Edwards
Low motor temperature - contact Edwards
A540-55-880 Issue D
Description
Page 61