7 – Service and Maintenance

Cleaning

Before connecting the pump to a patient, and periodically during use, clean the unit using a lint-free cloth
dampened with hydrogen peroxide 10% or 70% isopropyl alcohol.
Warning:
Warning:
Caution:
similar solvents. These chemicals can cause damage to plastic components. Use a lint-free
cloth dampened with hydrogen peroxide 10% or 70% isopropyl alcohol.
Caution:

Storage

If the pump is to be stored for an extended period, it should be cleaned and the battery fully
charged. Store in a clean, dry atmosphere at room temperature and, if available, employ the
original packaging for protection.
Perform functional tests and ensure that battery is fully charged once every three months.

Periodic Maintenance

Periodic maintenance (PM) is recommended every 1 year. The pump will display a
maintenance reminder alert for the user to send the pump for service yearly. The PM is
designed to assure the pump's accuracy and detect and repair any potential pump
inconsistencies prior to their occurrence in the field. During the PM, a biomedical engineer or
trained technician should perform the following procedures:
 Clean the pump thoroughly
 Visually inspect the pump to verify its structural integrity
 Perform all the manual tests in the Change Set Up menu
 Perform calibration procedures as per the Service Manual
 Run the pump for several hours to make sure no abnormalities occur during
infusion such as alarms, inaccurate infusion, and battery inconsistencies.
 NOTE: Service and maintenance should be performed by a certified CME technician. The
only maintenance a patient can perform is the changing of the battery and cleaning of the
device.
 NOTE: It is the CME technician's responsibility to repair any faults found during the
Periodic Maintenance.
 NOTE: Do not service the pump while it is in use and/or attached to patient.
BodyGuard 545 Color Vision™ Directions For Use
Always turn the pump off and remove the battery before cleaning.
Always unplug the charger from AC power before cleaning.
Do not clean the pump or charger with chemicals such as Xylene, Acetone or
Do not soak or immerse any part of the pump or charger in water.
7 – Service and Maintenance
71