Select Protocol (Start An Infusion) - CME BodyGuard 545 Color Vision Directions For Use Manual

Epidural infusion system
Hide thumbs Also See for BodyGuard 545 Color Vision:
Table of Contents

Advertisement

47. If the protocol parameters that were set
exceed the toxicity limit (MediGuard
the pump will not allow you to proceed to the
next step and will alarm 'MediGuard
Exceeded' and return to the 'Max
MediGuard
and amend your entry.
48. Finally confirm protocol parameters by
pressing
Or pressing
parameters.
NOTE: BodyGuard 545 Color Vision™ is capable of delivering the bolus
dose at up to 1200 ml/hr. The user can adjust bolus delivery down to a
minimum of 40 ml/hr in "Change set up mode" using the level three access
code.

Select Protocol (Start an Infusion)

Under normal, everyday circumstances most users of dedicated infusion pumps use the
pumps to perform the same or limited number of infusion protocols for the majority of
patients. For example, in pain management it is common for patients to be on a
morphine regime consisting of bolus only delivery of a concentration of 1-2mg/ml
morphine supplied by pharmacy in a fixed infusion bag volume. The service may use
different bolus volumes or lockout times depending on individual patient needs.
This means once you have set up these protocols, using the Level 2 Access Code in
'Modify Program Library' as described in the previous section, Level 1 code users can
access the pump and run the standard protocols for a patient without having to enter or
change any infusion parameters.
This is achieved as follows:
1.
Load the administration set as explained
previously. Turn the pump on by pressing
The following will appear.
BodyGuard 545 Color Vision™ Directions For Use
Rate' screen so you can check
TM
and going back to main menu,
to review and change
EPIDURAL
level)
TM
MEDIGUARD LIMIT
Limit
TM
EXCEEDED
EPIDURAL
A: Levobupivacaine
Confirm: Press
Review: Press
BodyGuard 545
.
Color Vision Epidural
Pump Serial Number:
Production Date:
Software Version
BGCV545_02B
Protocol
4 - Operation
10:00
-
70%
10:00
-
70%
V123456
1-Jan-2000
Sample
41

Advertisement

Table of Contents
loading

Table of Contents