Immunity Test
EMC Standard
Electrostatic
IEC 61000-4-2
Discharge (ESD)
Immunity
IEC 60601-2-24
Radiated RF
IEC 61000-4-3
Immunity
Proximity fields
IEC 61000-4-3
from RF wireless
communications
equipment
Electrical Fast
IEC 61000-4-4
Transients (EFT)
Immunity
Surge Immunity
IEC 61000-4-5
Conducted RF
IEC 61000-4-6
Disturbances
Immunity
Power
IEC 61000-4-8
Frequency
Magnetic Field
Immunity
Voltage Dips &
IEC 61000-4-11
Interruptions
BodyGuard 545 Color Vision™ Directions For Use
Test Level
Contact
±2, ±4, ±6 kV
Discharge
Air Discharge
±2, ±4, ±8
Contact
±8 kV
Discharge
Air Discharge
±15 kV
80 MHz - 2.7
10 V/m
GHz
380 – 390 MHz
27 V/m
430 – 470 MHz
28 V/m
704 – 787 MHz
9 V/m
800 – 960 MHz
28 V/m
1.7 – 1.99 GHz
28 V/m
2.4 – 2.57 GHz
28 V/m
5.1 – 5.8 GHz
9 V/m
±2 kV
±2 kV
Line-Line
± 0,5 kV, ± 1
kV
Line-Ground
± 0,5 kV, ± 1
kV, ± 2 kV
150 kHz - 80
3 V/m
MHz ISM and
6 V/m
Amateur Radio
bands
50 Hz & 60 Hz
30 A/m
0% for 0.5
0°, 45°, 90°,
cycle
135°, 180°,
225°, 270° and
315°
0% for 1 cycle
at 0°
70% for 0.5
at 0°
sec.
0% for 5 sec.
at any phase
angle
6 – Specifications
Compliance
Complies, No degradation
of performance
Complies, Temporary
degradation, loss of function
or performance which
requires operator
intervention is acceptable.
Complies, No degradation
of performance
Complies, No degradation
of performance
Complies, No degradation
of performance
Complies, No degradation
of performance
Complies, No degradation
of performance
Complies, No degradation
of performance
Complies, No degradation
of performance
70