Pump Accuracy - CME BodyGuard 545 Color Vision Directions For Use Manual

Epidural infusion system
Hide thumbs Also See for BodyGuard 545 Color Vision:
Table of Contents

Advertisement

Program Limits (including MediGuard
Under 'Change Set Up' users can choose from a number of options to limit protocol
parameters and set safe ceilings on drugs infused. When the MediGuard
is on users are asked to set the patients weight (kg) and a toxicity ceiling in ml, mg
or mcg per hour per kilo bodyweight. If users try to set a protocol where the
component elements (basal rate and boluses) exceeds the MediGuard
pump will alert the user to this and request they re-confirm their intentions, amend
or revise the toxicity ceiling. Level One users will not be able to change the protocol
or ceiling under Select Protocol and will have to select an alternative appropriate
protocol or consult senior clinical staff with Level 3 authority. Alternatively you can
choose to simply limit either the rate or number of boluses delivered over either a 1,
4, or 24 hour period according to local practice.

Pump Accuracy

The following graphs and curves were derived from testing described in IEC60601-2-24.
Testing was performed under normal conditions at room temperature (72° F or 22° C).
Any deviations from normal conditions and room temperature may cause changes in the
accuracy of the pump.
Start-up Curves
The Start-up curves represent continuous flow versus operating time for two hours from
the start of the infusion. They exhibit the delay in onset of delivery due to mechanical
compliance and provide a visual representation of uniformity. Trumpet curves are
derived from the second hour of this data. Tests performed according to IEC 60601-2-24
standard.
BodyGuard 545 Color Vision™ Directions For Use
TM
toxicity settings)
1 - Introduction
feature
TM
limit the
TM
13