Alarm Conditions And Troubleshooting; Alarm Condition; Alarm Definitions - CME BodyGuard 545 Color Vision Directions For Use Manual

Epidural infusion system
Hide thumbs Also See for BodyGuard 545 Color Vision:
Table of Contents

Advertisement

5 – Alarm Conditions and
Troubleshooting

Alarm Condition

When the infusion pump detects a problem, the following occurs:
The infusion stops
An audible/visual alarm is activated
A message appears on the display screen indicating the cause of the alarm
The LED indicator will change from green to red or yellow
NOTE: Pressing
NOTE: Operator's position to verify functionality of alarm system is one meter away from the
pump.
NOTE: Alarm settings are not lost in case of power loss. When power is lost for less than or
equal to 30 seconds, the alarm settings are restored.
NOTE: At Power-On the pump performs a sequence of audible beeps informing the user the
pump is ready for use. The beep sequence confirms self tests have verified the system
integrity. Continually, through pump operation the pump will perform a sequence of continuous
tests to verify system integrity. These self-tests check the system integrity against single fault
failures that may have occurred and will produce a System Error alarm if fault is detected.
Warning:
ineffective.
Warning:
or similar equipment in any single area.
Warning:
appropriate prior to use on each patient.

Alarm Definitions

Alerts and alarms are indicated by a combination of audible tones, visual flashing behavior, and a
descriptive message on the pump display.
Alarms notify the clinician and/or patient of a potential or an actual hazardous condition.
Terms
High Priority
BodyGuard 545 Color Vision™ Directions For Use
during an alarm mutes the alarm for two minutes.
Do not set alarm limits to extreme values. This may render the alarm system
A potential hazard may exist if different alarm pre-sets are used for the same
The operator should check to ensure that current alarm pre-sets are
Definitions
A high-priority alarm is initiated by any alarm condition requiring

5 - Alarm Conditions and Troubleshooting