7 – Service and Maintenance

Cleaning

Before connecting the pump to a patient, and periodically during use, clean the unit
using a lint-free cloth lightly dampened with warm water and a mild detergent or
disinfectant.
Warning: Always turn the pump off and remove the battery before cleaning.
Warning: Always unplug the charger from AC power before cleaning.
Caution: Do not clean the pump or charger with chemicals such as Xylene,
Acetone or similar solvents. These chemicals can cause damage to
plastic components and paint. Use a lint-free cloth dampened with
warm water and a mild detergent or disinfectant.
Caution: Do not soak or immerse any part of the pump or charger in water.

Storage

If the pump is to be stored for an extended period, it should be cleaned and the battery
fully charged. Store in a clean, dry atmosphere at room temperature and, if available,
employ the original packaging for protection.
Perform functional tests and ensure that battery is fully charged once every three
months.

Battery Operation

The BodyGuard® pump can operate on battery power which enables operation when
the patient is being moved or during electrical power failure. When the pump operates
on battery power, the AC icon is off. At full charge, the standard battery provides 15-17
hours of operation at an infusion rate of 125 ml/hr.
Warning: Do not operate the pump on AC power if the battery is not loaded in
the pump for back up.
Note:
When the pump is not in use, click the pump into the Charger and
plug the system into an AC wall outlet (if possible) to charge battery.
Software ver. CVPCA30E_1