The BodyValve
The Anti-Free Flow valve, BodyValve
Preventing free-flow in the event the set is detached from the pump.
Preventing back-flow (reflux) in the event several infusion pumps are connected
simultaneously to the same patient.
Preventing free-flow in the event of a mechanical malfunction.
Preventing pump operation if the set has been loaded incorrectly.
Warning: Replace the administration set every 72 hours to lessen the
incidence of bacteria formation.
Warning: Injury may occur to patient or operator injury if package is opened
or damaged, or if damaged components are used. Visually inspect contents
and package before each use.
Warning: Air embolization can cause death or serious injury to the patient. Do
not connect a patient to the BodyGuard 545 Color Vision
all trapped air has been cleared from the fluid path. Carefully read the
instructions for loading. Puncturing set components may cause air embolism.
Warning: Drugs must not be administered to the epidural space unless the
drugs are indicated for this purpose and are administered in accordance with
the indications included in the manufacture's package insert accompanying the
drugs. Epidural administration of drugs other than those indicated for epidural
use could result in serious injury to the patient.
Caution: Component damage may occur if not installed properly. Assure all
connections are secure; do not over-tighten. This will help minimize leaks,
disconnection, and component damage.
BodyGuard 545 Color Vision™ Directions For Use
TM
, enhances pump functioning by:
TM
4 - Operation
Infusion Pump until
TM
26