NOTE: If using an ambulatory administration set without a drip chamber, priming
the line might be needed. Disconnect the set from patient before priming.

Eliminating Air During Operation

When air is detected in the line, the pump stops the current infusion, sounds an audible
alarm, the screen shows Air or Up Occlusion, and the LED indicator turns from green to
red.
To eliminate air in the line during pump operation, follow these steps:
Press
to silence the alarm. Press and hold down
1.
the menu screen.
Enter the Access Code.
2.
Select Prime and press
3.
4.
Enter the relevant Access Code.
5.
Disconnect the line from the patient; use aseptic technique.
Warning: ensure the line is disconnected from
the patient before proceeding to the next step.
Failure to do so could result in the patient
receiving air through into the bloodstream.
Press
within 10 seconds to commence priming, otherwise the display will back
6.
1 step automatically. You may cease priming at any time by pressing
7.
Repeat steps 3-6 until all air is eliminated from the line and fluid is seen at the
distal end of the set.
Press
to return to the menu option screen.
8.
Press
to resume the current infusion.
9.
10.
Check the Infusion Summary if screen which shows: bag volume, volume left and
infused, is correct then press
BodyGuard 545 Color Vision™ Directions For Use
.
again to restart delivery.
for 3 seconds to return to
EPIDURAL
10:00
Warning!
Disconnect Patient
Press
key to Prime
4 - Operation
-
70%
.
55