7 – Service and Maintenance
Cleaning
Before connecting the pump to a patient, and periodically during use, clean the unit using a
lint-free cloth lightly dampened with warm water and a mild detergent or disinfectant.
Warning: always turn the pump off and remove the battery before cleaning.
Warning: always unplug the charger from AC power before cleaning.
Caution: do not clean the pump or charger with chemicals such as Xylene,
Acetone or similar solvents. These chemicals can cause damage to plastic
components and paint. Use a lint-free cloth dampened with warm water and a
mild detergent or disinfectant.
Caution: do not soak or immerse any part of the pump or charger in water.
Storage
If the pump is to be stored for an extended period, it should be cleaned and the battery
fully charged. Store in a clean, dry atmosphere at room temperature and, if available,
employ the original packaging for protection.
Perform functional tests and ensure that battery is fully charged once every three months.
Battery Operation
The BodyGuard 595 Color Vision™ pump can operate on battery power which enables
operation when the patient is being moved or during electrical power failure. When the
pump operates on battery power, the AC icon is off. At full charge, the standard battery
provides 17 hours of operation at an infusion rate of 125 ml/hr.
Warning: do not operate the pump on AC power if the battery is not loaded in
the pump for back up.
NOTE: when the pump is not in use, click the pump into the Charger and plug
the system into an AC wall outlet (if possible) to charge battery.
NOTE: after the "End Battery" signal has been activated or following long
periods of storage, wait 2 minutes after the pump has been connected to an
AC power supply before operating.
Caution: leaving the battery in an uncharged state for a long period of time
may damage the battery.
BodyGuard 595™Color Vision Operator Manual
7 – Servi c e and Maintenance