2 – Symbols, Warnings and Cautions
Infusion of blood or blood products should be retricted to a maximum rate of 600 ml/h.
Cautions
To avoid possible damage to the equipment, observe the following:
Leaving the battery in a depleted state for a long period of time may damage the battery.
Connect the pump to the mains via the charger whenever possible to recharge the battery.
Do not clean the pump or charger with chemicals such as Xylene, Acetone or similar
solvents.These chemicals can cause damage to plastic components and paint. Use a lint-free
cloth with hydrogen peroxide 10% or isopropyl alcohol 70%.
Immersing the BodyGuard 545 Color Vision™ infusion pump into liquid could cause damage to
components. Do not immerse the pump into any type of liquid.
The BodyGuard 545 Color Vision™ infusion pump is not certified for use in oxygen-enriched
environments.
Battery damage could occur if left in a temperature warmer than 50C (122F).
Potential strangulation may occur if the cables/tubing are of excessive length.
Potential choking may occur if small parts are inhaled or swallowed.
Potential allergic reactions may occur due to materials used in the pump.
The front panel push buttons may become hot to the touch when the pump is operated in an
elevated temperature environment up to 45 °C (i.e., outdoors or in a hot automobile).