• This guide is not intended to be comprehensive instructions for the setup and operation of the infusion pump. For complete pump information, refer to the BodyGuard 323 Color Vision™ Directions For Use . • Some screens in this guide may vary depending on the way the pump is set up.
Page 2
Loading Infusion Set • Ensure the pump is secured in the pump charger. STEP 1 Ensure the pump and charger are fastened to the pole (if using pole) using the roller wheel at the back of pump charger. BodyGuard Color Vision 323 • Switch on the pump by pressing .
Page 3
Loading Infusion Set STEP 4 • Insert the bag spike into the fluid container. • If a drip chamber is present, fill the drip chamber halfway. STEP 5 • Pull down the door latch to open the door. • Securely fit the black key into the pressure sensor fitment on the right side. • The two protrusions must face up. The key can only be loaded in this direction. • Ensure that the flow direction is in line with the flow direction arrows inside the pump.
Page 4
Infusion Set Priming Continuous, 25-Steps, TPN, Intermittent, Drug Library STEP 1 • To prime and eliminate the air from the set, press key. Continuous 10:00 ███■ • Disconnect the patient prior to starting the prime Prime function Warning ! Disconnect Patient Press key to Start The prime function must not be performed while the IV set is...
Page 5
Start New Infusion Continuous STEP 1 • Select mL/h or press the key to switch to mg/kg units. Pressing the key again will Continuous 10:00 ███■ switch to mcg/kg units. mL/hr • To switch from mg/mcg units to ml/h, enter a Press Key to Continue value of “0”...
Page 6
Start New Infusion Continuous STEP 4 • Review the infusion parameters and confirm. • To make changes, press the key to return Continuous 10:00 ███■ back. Rate 125 mL/hr Program Volume 1000 mL Volume Left 1000 mL Volume Infused 0 mL Air Sensor 0.3 mL Pressure 730 mmHg Press...
Page 7
Rate Titration, Keypad Lock, History Rate Titration To change rate during infusion: • Enter a new rate and confirm with key. Continuous 10:00 ███■ Rate • Single alarm sounds as confirmation. Make sure the new rate is displayed on screen. mL/hr ███████████ █ ████████████████ mmHg Titration may not be available in all modes or setups. Keypad Lock Keypad lock: prevents unauthorized changes: • Press and hold the...
Page 8
Secondary (Piggyback) / Bolus Infusion Continuous To deliver a secondary (Piggyback) infusion or a STEP 1 bolus: Continuous 10:00 ███■ Continuous 10:00 ███■ • Press the key. Secondary Infusion Piggy Rate Bolus Dose • Enter the desired rate/ dose. mL/hr Adjust and press Adjust and press Go to pump settings to select either Piggyback or bolus system.
Page 9
Start New Infusion 25 Steps 25-Steps 10:00 ███■ Rate (Step 1) 25 – Steps VAL ml/h Program Press & Hold to change Volume (Step 1) VAL minutes Rate (Step 2) VAL ml/h Volume (Step 2) VAL minutes When Done Enter “0”, Press Review Summary...
Page 10
Start New Infusion 10:00 ███■ Bag Volume VALUE mL Program Press & Hold to change Volume TBI VALUE mL Total Time HH:mm Up Time HH:mm Down Time HH:mm KVO Rate 0-20 mL/hr Review Summary Start Infusion? For Early Taper Down, press and hold key.
Page 11
Start New Infusion Intermittent Intermittent 10:00 ███■ Bag Volume Intermittent VALUE mL Program Press & Hold to change Dose Volume VALUE mL Number of Doses Minimum Dose Rate VALUE mL/hr Dose Duration HH:mm Interval Time HH:mm KVO Rate 0-20 mL/hr Review Summary Start...
Page 12
Start New Infusion 10:00 ███■ Program Key to Continue Press 10:00 ███■ 10:00 ███■ Prime Prime Program Infusion Program Infusion Clinician Bolus Clinician Bolus Patient History Patient History Review Set up Review Set up Change Set up Change Set up Select and press Select...
Page 13
Start New Infusion Set VALUE Basal Rate to Zero for VALUE ml/hr, mg/hr, mcg/hr Bolus Only Maximum Titration Rate VALUE ml/hr, mg/hr, mcg/hr Bolus Dose VALUE ml, mg, mcg Maximum Bolus Dose VALUE ml, mg, mcg Lock out Time VALUE minutes Program Time 0-24 Volume Limits VALUE...
Page 14
Start New Infusion Drug Library Drug Library 10:00 ███■ Select Drug Drug Library Program Library Demo DEP1 Department Patient Weight VALUE kg Dose VALUE mg/hr, mg/min, mcg/hr, mcg/min, mcg/kg/min, mg/kh/hr mcg/kg/hr, mg/kg/min Volume VALUE Review Summary Start Infusion? 14/24 | QRG 999-603EN Rev. 00...
Clinician Bolus Press Key During Infusion Press Bolus Key Code XXX Clinician Bolus Dose VALUE Bolus Infusion Bolus Delivery Drug Library Press Key During Infusion Bolus Volume Bolus Weight per Body Weight VALUE mL VALUE mg, mcg, mg/kg, mcg/kg (Optional) Bolus Duration Can be Set VALUE...
Alerts & Alarms Resume/Repeat • Pump was switched off before the last infusion has ended. ◊ Select Resume to continue from the last Continuous 10:00 ███■ stopped position. Resume – Press Repeat – Press ◊ Select Repeat to start a new infusion. End of Infusion • The infusion has ended.
Page 17
Alarms Pump Paused • Pump “Stopped” or unattended for 2 minutes or more. Resume infusion or turn off the pump. Continuous 10:00 ███■ • If the pump is turned off, you will have the option Pump Paused to resume from the last position. Too Long Press Missing Key...
Page 18
Alarms High Flow • High Priority Alarm. Infusion stops. • High drops counted by drop sensor. • Check the IV bag and drop sensor position. • Drop sensor is experiencing high drops. • If alarm is not resolved, send the pump for service. • The pump is set up to work with a drop sensor.
Cleaning & Maintenance Before connecting the pump to a patient, and periodically during use, clean the unit as instructed below. • Always turn the pump off and remove the battery before cleaning. • Always unplug the charger from AC power before cleaning. • Do not clean the pump or charger with chemicals such as Xylene, Acetone, or similar solvents.
Intended Use The BodyGuard 323 Color Vision™ infusion pump system is designed for infusion of medications or fluids requiring continuous or intermittent delivery at precisely-controlled infusion rates through clinically acceptable routes of administration, including intravenous, subcutaneous, percutaneous, in- tra-arterial, epidural, intrathecal, in close proximity to nerves, and into an intraoperative site (soft tissue/body cavity/surgical wound site).