43. Set KVO rate and press
applicable.
NOTE
If no KVO is required, set value to "0".
:
NOTE: If KVO is disabled in Change Set Up Menu, the KVO screen will not be
displayed.
.
44. Set occlusion pressure alarm level specific to
the protocol/patient and press
45. Set air-in-line limit specific to the
protocol/patient and press
46.
If the MediGuard
pump will ask for 'Max MediGuard
ml, mg or mcg per hour per kg bodyweight.
This sets a toxicity limit for the protocol based
on the patient's weight.
NOTE
If MediGuard
:
Change Set Up menu this screen will not be
displayed.
BodyGuard 545 Color Vision™ Directions For Use
. The to 0 if not
.
.
feature is enabled, the
TM
TM
is set to OFF in the
TM
EPIDURAL
A: Levobupivacaine
Keep Vein Open Rate
0
.
5
Enter and Press
EPIDURAL
A: Levobupivacaine
Occlusion Pressure
350
(250-1100 mmHg)
Enter and Press
EPIDURAL
A: Levobupivacaine
Air-in-line Limit
0
.
1
Enter and Press
EPIDURAL
Rate' in
A: Levobupivacaine
MediGuard
On
Select
EPIDURAL
A: Levobupivacaine
Max MediGuard Rate
8
mg/kg/hr
Enter and Press
4 - Operation
10:00
-
70%
mL/hr
Key
10:00
-
70%
mmHg
Key
10:00
-
70%
mL
Key
10:00
-
70%
,
and press
10:00
-
70%
Key
40

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the BodyGuard 545 Color Vision and is the answer not in the manual?

Table of Contents