Pump Parameters

The pump comes programmed with a number of parameters that can be viewed using
'View Set Up' from the main menu and changed using the technician Access Code in
'Change Set Up' mode.
Parameter
Battery Level
Time, Date
Keystroke Volume
Alarm Sound Volume
Priming Volume
Priming Rate
Maximum Bolus
Maximum CA Bolus
Maximum Bolus Rate
Maximal Basal Rate
Keep Vein Open Rate
Program Limits
Minimum Lockout
Bolus Titration
Info screen duration
Language
Serial Number
Production Date
Software Version
Main Self-test
Volume Test
Manual Test
BodyGuard 545 Color Vision™ Directions For Use
Purpose
Current battery charge
Current time and date
Volume of keystroke
Volume of alarm
(on view menu can be found under
keystroke volume)
Amount of fluid delivered during
set priming
Rate at which priming occurs.
Maximum bolus volume for a
protocol
Maximum bolus volume allowed by
a clinician
Rate at which boluses are
delivered
Maximum continuous basal rate
A continuous rate after infusion
program has ended which is
designed to prevent coagulation
and keep an IV line open
MediGuard
Minimum lockout time between
boluses
Change bolus volume during
infusion
Number of seconds the info screen
is displayed before returning to the
current infusion screen
Change between available
language (may vary between
software version)
Pump unique serial number
Date of manufacture
Pump software version
Automatically test all pumps
hardware features
Test the volume calibration
Manually test the pumps hardware
5 - Alarm Conditions and Troubleshooting
Range
Empty >> Full
0-7
4-6
0.1-20ml
300-1000 ml/hr
20ml
0.1-60ml
40-1200ml
System will adapt rate
according with in-line
backpressure
0.1-30 ml/hr
0 ml/hr (Disabled)
0.1-5.0ml/h
ON/OFF
0-240min
ON/OFF
1-255 sec
English, Spanish, Italian,
German (Deutch), French,
Dutch
Fixed data
Day-month-year
Fixed data
prime volume, continuous
rate, bolus dose, bolus
rate, bolus lock-out
Keypad test, display test,
63