Electromagnetic Immunity - Nuvo Navigator Lift 150 Single Installation Instructions Manual

Table of Contents

Advertisement

19
Electromagnetic Compatibility (EMC) Information
Interference immunity test
Electrostatic discharge in accordance with
IEC 61000-4-2
Electrical fast transients / bursts in accor-
dance with
IEC 61000-4-4
Surges
Line-to-line in accordance with
IEC 61000-4-5
Surges
Line-to-earth in accordance with
IEC 61000-4-5
Voltage dips in accordance with
IEC 61000-4-11
Voltage interruptions in accordance with
IEC 61000-4-11
Power frequency magnetic field immunity in
accordance with
IEC 61000-4-8
1568911, Edition 2019-04, Version 0

19.1.2 Electromagnetic immunity

The Navigator Lift
150 / 250 is intended for use in the ELECTROMAGNETIC environ-
ment specified below. The customer or the user of the Navigator Lift
sure that it is used in such an environment.
Test level in accordance with IEC 60601
±8kV contact
±2kV, ±4kV, ±8kV, ±15kV air
±2kV
100kHz repetition rate
±0.5kV, ±1kV
±0.5kV, ±1kV, ±2kV
0% U
; 1/2 period
T
at 0, 45, 90, 135, 180, 225, 270 and 315 degrees
0% U
; 1 period
T
and
70% U
; 25/30 periods
T
Single phase: at 0 degrees
0% U
; 250/300 periods
T
30A/m
50Hz or 60Hz
Please note:
U
is the a.c. mains voltage prior to application of the test level.
T
88
150 / 250 should en-
Test result
passed
passed
passed
passed
passed
passed
passed
passed
EN