18
Technical Data
Maximum loading capacity
on the pendant system
Protection class / type
Medical Device Directive 93/42/EEC
Applicable standards,
laws and directives
Approvals of the standard equipment
Motor arm 150 (1015mm long) .................................................................................150kg
Motor arm 250 (1015mm long) .................................................................................250kg
Extension arm, 600mm, with motor arm 150 (1015mm long) ..................................150kg
Extension arm, 800mm, with motor arm 150 (1015mm long) ..................................150kg
Extension arm, 1000mm, with motor arm 150 (1015mm long) ................................150kg
Extension arm, 1200mm, with motor arm 150 (1015mm long) ................................140kg
XL extension arm, 600mm, with motor arm 250 (1015mm long) .............................250kg
XL extension arm, 800mm with motor arm 250 (1015mm long) ..............................250kg
XL extension arm, 1000mm with motor arm 250 (1015mm long) ............................250kg
XL extension arm, 1200mm with motor arm 250 (1015mm long) ............................250kg
XL extension arm, 1400mm with motor arm 250 (1015mm long) ............................240kg
XL extension arm, 1600mm with motor arm 250 (1015mm long) ............................200kg
Electrical data
Rated voltage.............................................................AC 100V-120V and AC 220V-240V
Rated frequency ................................................................................................. 60 / 50Hz
Rated voltage at AC 100V-120V ................................................................................ 10A
Rated voltage at AC 220V-240V .................................................................................. 5A
Indirect extension arm lighting ............................................................................... DC 12V
2 / 4 lighting boards (supply voltage 12V DC, 2 lighting boards each connected in series
to 24V DC)
Noise level
Sound energy level .................................................65db(A) (EN ISO 3746) not exceeded
............................................................................................................................................
Brake torque
Brake torque with the pneumatic brake actuated ....................................... approx. 50Nm
Brake torque with the electromagnetic brake actuated on the motor arm ... approx. 70Nm
Brake torque with the electromagnetic brake actuated on the
extension arm and motor arm ............................................ approx. 70Nm / approx. 70Nm
Brake torque with the electromagnetic brake actuated on the
XL extension arm and motor arm .................................... approx. 150Nm / approx. 70Nm
Manual forces
Dynamic torque with the electromagnetic brake released depending on the position and
payload ........................................................................................................... 3.5 to 40Nm
Protection class in accordance with IEC 60601-1..............................................................I
IP classification in accordance with IEC 60529 ........................................................ IP 20
Classification......................................................................................................................I
Medical Devices Act (MPG)
93/42 EEC (Medical Device Directive)
IEC 60601-1 – Medical Electrical Equipment – Part 1: General Requirements for Basic
Safety and Essential Performance.
Recognised NRTL component.