Ambient Conditions; Ambient Conditions For Storage And Transport; Ambient Conditions For Operation; Information On The Rating Plate - Nuvo Navigator Lift 150 Single Installation Instructions Manual

Table of Contents

Advertisement

2
Safety Instructions
Figure 5: Information on the rating plate
TYPE:
Navigator Lift,250,Single
Extension Arm 1000 mm
AC 100-120V/50-60Hz/max. 10A
AC 220-240V/50-60Hz/max. 5A
S2- 3min (30min)
max. load 250 kg
Manufactured for
Medical Illumination International, 547 Library Street, San Fernando, CA 91340 USA
2.7

2.7.1 Ambient conditions for storage and transport

The following conditions apply to storage:

2.7.2 Ambient conditions for operation

2.8
(See "Figure 5")
6000000
WWYYW000000000
2019
Serial number
Power supply
S2-3min (30min) duty cycle of the height adjustment mechanism
Load bearing capacity
Date of manufacture
19

Ambient conditions

Ambient temperature:
Relative humidity:
Atmospheric pressure:
Store only in indoor rooms.
Ambient temperature:
Relative humidity:
Atmospheric pressure:
This corresponds to a maximum operating altitude of 3,000m.

Information on the rating plate

The rating plates are attached to the top side of the extension arm or
XL extension arm and under the front cover of the motor arm.
Navigator Lift
150 / 250 is the version equipped with electromag-
netic brakes.
Navigator Lift
XL 250 is the version equipped with the XL extension
arm.
The rating plate indicates the serial number (SN) of the pendant sys-
tem.
The rating plate provides information on the power supply of the pen-
dant system.
See Chapter 18, "Technical Data", on page 84.
Max. load indicates the maximum permissible carrying load of the
motor arm.
The max. load of the Navigator Lift
The max. load of the Navigator Lift
The digits 1 to 4 of the serial number (SN) indicate the date of man-
ufacture of the pendant system.
The first two digits indicate the week of manufacture, e.g. 14 = cal-
endar week 14.
The following two digits indicate the year of manufacture, e.g.
15 = 2015.
The letter in the 5th position indicates the factory, e.g. H = Hünfeld.
The digits following the letter indicate the serial number.
The information and illustrations serve as examples.
The information and illustrations on the rating plate can vary.
-25 °C to 70 °C
10 % to 75 %
500 hPa to 1,060 hPa
10 °C to 40 °C
30 % to 75 %
700 hPa to 1,060 hPa
150 is 150kg
250 is 250kg
1568911, Edition 2019-04, Version 0