12
Connecting the Supply Cables and Hoses
Figure 39: Installing gas supply hoses and exhaust air ducts
Interface plate
anaethesia
A - ir
gas
motor
evacuatio
bar
bar
bar
0
Indicator
Navigator M6
Gas supply points
bar
bar
bar
N
0
Va uum
c
Air
2
2
Manometer
12.2 Installing gas supply hoses and exhaust air
ducts
(See "Figure 39")
1.
Observe the safety instructions in Chapter 12.1 on Page 60.
NOTE – Make sure that the gas types are correctly assigned
The gas type is indicated on a label attached to the gas supply hoses
equipped with sealing plugs.
The sealing plugs may only be removed during installation.
Check the hoses and ducts for contamination and clean them with
oil-free air.
The illustration shows the example of a gas supply hose. Make sure that
the cables, hoses and ducts are assigned to the correct supply outlet
points.
1.
Check the gas supply hoses for contamination and clean them with
oil-free air.
2.
Fit a hose clamp to the gas supply hose, remove the sealing plug
and push the hose onto the correct gas supply outlet point.
Up to 3 gas outlets and up to 2 vacuum outlets may be connected to
a gas valve using Y connectors.
3.
Press on the hose clamp and check that it is securely in place.
4.
Connect and secure the anaesthetic gas suction hoses and the air
motor exhaust air ducts.
5.
Perform a gas type test as described in Chapter 13.1 on Page 65.