Electrical Connections: General Instructions - Xylem DIQ/S 281 Operating Manual

Universal transmitter for one digital sensor
Table of Contents

Advertisement

DIQ/S 281
Cable glands
ba76226e02
6
Fix the impact protection plate (Pos. 4) to the back of the DIQ/S 281 with
the bolts (Pos. 5).
3.7

Electrical connections: General instructions

All electric cables are fed from below through openings prepared in the enclo-
sure of the DIQ/S 281 and the DIQ modules. Cable glands with different clamp-
ing ranges are included with the DIQ/S 281 to provide sealing between the cable
and enclosure as well as for strain relief. Select the suitable cable gland for the
relevant cable type and cable diameter:
 Small, clamping range 4.5 to 10 mm.
This cable gland is suitable for all cables (earth cable after stripping the outer
insulation, see section 3.6) and sensor connection cables.
 Large, clamping range 7 to 13 mm.
This cable gland is required for cable sheaths with an outside diameter of
more than 10 mm and is screwed into the enclosure via an extension piece.
If necessary, you can order more big cable glands (see
chapter 10 A
03/2023
sealing ring 20 x 15 x 1 mm
cable gland M16
blind plug
sealing ring 20 x 15 x 1 mm
extension piece M16/M20
sealing ring 24 x 19 x 2 mm
cable gland M20
CCESSORIES AND OPTIONS
).
Installation
41

Advertisement

Table of Contents
loading

This manual is also suitable for:

Wtw diq/s 281

Table of Contents