P R E L I M I N A R Y
D
5813707-100, R
IRECTION
8-2-19
Printer and Printer Bracket Replacement Procedure
8-2-19-1
Tools
Phillips screwdriver of appropriate size
FRU Part #
Refer to
8-2-19-2
Time Required
5 min
8-2-19-3
Preparations
1.) Shut down the Venue Go™ ultrasound unit, as described in
2.) Prepare stable flat surface covered with a soft material (such as foam) to prevent scratches to the
display.
3.) Dismount Venue Go™ system from cart and place it on a previously prepared surface with the
display facing down.
8-2-19-4 Printer and Printer Bracket Removal Procedure
4.) Remove printer assembly
•
5.) Remove 4 Phillips screws holding the printer inside the printer tray.
8-48
2
EVISION
Table 9-16
.
on page 9-17
Printer Assembly Removal Procedure
Figure 8-128 Disconnect Printer Screws
Chapter 8 - Replacement Procedures
V
™ S
ENUE
ERVICE
Power Shutdown
.
M
ANUAL