Safety Checks

4-3-2-1
Ground Continuity Check
1) Use an Ohmmeter/multimeter to measure the ground resistance between the Split cable GND pin
and the cart GND pin.
2) Verify the measured value is less than 0.1 [Ohm].
4-3-3

Phantoms Performance Checks

The use of test phantoms is only recommended if required by your facility's (customer's) QA program.
4-10
2
EVISION
Figure 4-48 GND Continuity Test