Casters Replacement Procedure - GE Venue Go Service Manual

Table of Contents

Advertisement

P R E L I M I N A R Y
D
5813707-100, R
IRECTION
8-2-17

Casters Replacement Procedure.

8-2-17-1
Tools
Allen key 6 mm
Phillips screwdriver of appropriate size
FRU Part #
Table 9-16
Refer to
8-2-17-2
Time Required
5 min
8-2-17-3
Preparations
1.) Shut down the Venue Go™ ultrasound unit, as described in
2.) Prepare stable flat surface covered with a soft material (such as foam) to prevent scratches to the
display.
3.) Dismount Venue Go™ system from cart and place it on a previously prepared surface with the
display facing down.
4.) Lower the cart tray to its lowest position.
8-2-17-4
Casters Removal Procedure
5.) Using a flat screw driver, carefully remove the screw cap.
6.) Using Allen key 6mm, remove the holding screw and detach the caster.
8-2-17-5
Casters Installation Procedure
1.) Install new caster and secure the holding screw.
2.) Place screw cap back into its place.
3.) Mount Venue Go™ system on cart.
2
EVISION
.
on page 9-17
Figure 8-125 Removing Cater Cap and Holding Screw
Chapter 8 - Replacement Procedures
V
™ S
ENUE
ERVICE
Power Shutdown
.
M
ANUAL
8-45