Printer Assembly Replacement Procedure

8-2-18-1
Tools
Phillips screwdriver of appropriate size
FRU Part #
Refer to
8-2-18-2
Time Required
5 min
8-2-18-3
Preparations
1.) Shut down the Venue Go™ ultrasound unit, as described in
2.) Prepare stable flat surface covered with a soft material (such as foam) to prevent scratches to the
display.
3.) Dismount Venue Go™ system from cart and place it on a previously prepared surface with the
display facing down.
Printer Assembly Removal Procedure
8-2-18-4
4.) Disconnect printer cables: USB, printer power supply.
Disconnect printer PWR supply cable
8-46
2
EVISION
Table 9-16
.
on page 9-17
Figure 8-126 Disconnect printer USB and Power Cables
Chapter 8 - Replacement Procedures
V
™ S
ENUE
ERVICE
Power Shutdown
.
Disconnect
printer USB
Disconnect
printer PWR
supply
M