C-Psb Replacement Procedure - GE Venue Go Service Manual

Table of Contents

Advertisement

P R E L I M I N A R Y
D
5813707-100, R
IRECTION
8-2-4

C-PSB Replacement Procedure

8-2-4-1
Tools
Phillips screwdrivers of appropriate size
FRU Part #
Refer to
8-2-4-2
Time Required
5 min
8-2-4-3
Preparations
1.) Shut down the Venue Go™ ultrasound unit, as described in
2.) Prepare stable flat surface covered with a soft material (such as foam or thick cloth) to prevent
scratches to the display.
3.) Dismount Venue Go™ system from cart as described in
and place it on a previously prepared surface with the display facing down.
8-2-4-4
C-PSB Removal Procedure
1) Remove back cover.
2) Set the lower lever upwards to expose the screw, and remove four screws that hold the board.
8-8
2
EVISION
Table 9-16
.
on page 9-17
Back Cover Removal Procedure
Figure 8-78 Removing Four PSB Screws