GE Venue Go Service Manual page 218

Table of Contents

Advertisement

P R E L I M I N A R Y
D
5813707-100, R
IRECTION
5) Slip the broken probe lever out;
8-2-5-5
Probe Lever Installation Procedure
6) Attach the black rubber seal to the lever. Then slip the lever into the axle.
Note:
The lever has a longer (marked in Blue) and shorter arc. Ensure the lever is inserted with
its long arc as shown in the picture below:
7.) Slip the black plastic cap to secure the lever on the axle.
8.) Attach 2 screws holding the probe lever cap.
9.) Test the functionality of the probe locking before proceeding.
10) Install C-PSB.
11.)Install back cover.
8-12
2
EVISION
Figure 8-83 Slip Probe Lever Out and remove rubber seal
Figure 8-84 Insert Black Rubber Seal and Place Lever on the Axle
C-PSB Installation Procedure
Back Cover Installation Procedure