The Battery Temperature sensor signal is connected to its charger.
The Charger has a sensor signal connected to it, and it automatically stops charging in case the battery
temperature exceeds 46±2° C.
The Charger resumes normal charging when the battery temperature is at least below 43° C.
5-9-3-3
Battery Power Specifications
Scanning time in continuous 2D mode with new and fully charged pair of batteries, using12Ln-RS probe
with "vascular" factory preset is MIN: 120 minutes.
Scan time in 2D after 24 hrs of standby mode from new, fully charged batteries is MIN: 90.0 min.
Scan time in 2D after 48 hrs of standby mode from new, fully charged batteries is MIN: 60 min.
The system can remain in standby mode for 3 days (72.0 hours), using a fully charged, new battery.
Chapter 5 - Venue Go™Components and Function (Theory)