P R E L I M I N A R Y
D
5813707-100, R
IRECTION
8-2-20

Cart Tray Replacement Procedure

8-2-20-1
Tools
Phillips screwdriver of appropriate size
FRU Part #
Refer to
8-2-20-2
Time Required
5 min
8-2-20-3
Preparations
1.) Shut down the Venue Go™ ultrasound unit, as described in
2.) Prepare stable flat surface covered with a soft material (such as foam) to prevent scratches to the
display.
3.) Dismount Venue Go™ system from cart and place it on a previously prepared surface with the
display facing down.
8-2-20-4
Cart Tray Removal Procedure
4.) Dismount the system from the cart.
5.) Remove the cradle assembly.
6.) Remove the tilt assembly.
7.) Remove the cart bin.
8.) Remove the printer assembly.
8-50
2
EVISION
Table 9-16
.
on page 9-17
Dismounting System on Cart
Cradle Assembly Removal Procedure
Cradle Tilt Removal Procedure
Cart Bin Removal Procedure
Printer Assembly Removal Procedure