GE Venue Go Service Manual page 67

Table of Contents

Advertisement

P R E L I M I N A R Y
D
5813707-100, R
IRECTION
2-3-2-1
Suggested Minimal Floor Plan
NOTE:
GE requires a dedicated power and ground for the proper operation of its Ultrasound equipment. This
dedicated power shall originate at the last distribution panel before the Ultrasound system.
The Ultrasound system will function on voltages from 100-240 Volts and 50 or 60 Hz. However, if using
220 volt power in North America, then a center tapped power source is required.
Sites with a mains power system with defined Neutral and Live:
The dedicated line shall consist of one phase, a neutral (not shared with any other circuit), and a full
size ground wire from the distribution panel to the Ultrasound outlet.
Sites with a mains power system without a defined Neutral:
The dedicated line shall consist of one phase (two lines), not shared with any other circuit, and a full
size ground wire from the distribution panel to the Ultrasound outlet.
1.
Door – at least 762 mm (30 inches)
2.
Film Viewer
3.
Counter Top, Sink with hot and cold water and
Supplies Storage
4.
Linen Supply
5.
Probes/Supplies
6.
Examination Table – 1930 x 610 mm (76 x 24
inches)
7.
Footswitch
2
EVISION
Scale:
Each square equals one square foot (app. 31 x 31 cm)
Figure 2-1 Minimal floor plan, 2.5 m x 3 m (8 by 10 foot)
Chapter 2 - Site Preparations
8.
Stool
9.
Ultrasound system
10.
External Peripherals
11.
Dedicated Power Outlet -
Circuit Breaker protected and easily accessible
12.
Network Interface
13.
457 mm (18 inches) distance of Ultrasound
system from wall or objects
14.
GE Cabinet for Software and Manuals
V
G
™ S
M
ENUE
O
ERVICE
ANUAL
2-9